Low Volume Aerosol Sampler DPA-14

with automatic filter changer for autonomous, continuous aerosol sampling according to the standard EN12341

 

Features

  • Touchscreen interface
  • Automatic filter change
  • Change failure recognition
  • Empty magazine recognition
  • Overload cut-off
  • Internal data memory
  • Interchangeable filter magazines
  • Valve and software for easy tightness test and checks according to EN:12341
  • Venturi type orifice, P/T compensated
  • Ethernet port for remote control and data query
  • Field housing: Weather-proof housing made of aluminium, protected with an extremely weather and seawater resistant powder coating
  • Remote control and filter list upload
  • EN12341: 2014 compliance verified by TüV, EN12341: 2023 type testing in progress
  • TSP, PM10, PM2.5, PM1 inlets

Description

DIGITEL Low Volume Samplers DPA-14 are fully automatic systems to sample dust and aerosol particles for later assessment and analysis (gravimetric and analytical determination) in accordance with EN12341. The sampler operation range in standard execution is 15 to 50 litres per minute (0.3 to 3m³/h). The DIGITEL LVS DPA-14 has a magazine of 30 filters stretched in filter holders. They are automatically changed to the flow position at the pre-set time. The devices can be integrated in automatic monitoring systems via various interfaces. The field housing of the DIGITEL LVS DPA-14 is suited for outdoor installation. It is easy to transport and because of a good sound insulation very quiet. Superior workmanship in sampler mechanics backed by the latest technical and electronic control guarantee a long lifetime and absolutely reliable operation.

Specifications

Sampling
  • Flow rate: 15 – 50 l/min
  • Filter: round filters of d = 47 mm, flowing area d = 40 mm
  • Filter changer capacity: 30 filter in filter magazine plus one in sampling position
  • Application range: -20° to 50° C; 0 % to 95 % RH, -40° to 50° C with additional heater, 2000 amsl
Time programs
  • Work, Pause (0 to 99’999 minutes each)
  • Start time adjustable, using date and time
  • Different sampling cycles programmable
Protocol files
  • Data of filter, temperature, pressure, humidity, flow, blower load
  • Calibration history, sensor mean value record file
  • Settable averaging period 1 min – 24 h
Accuracy
  • Constancy of sample flow: <2% with calibration at 20°C, operating at -20°C-50°C
Operating Data
  • Power supply: 230V AC / 50-60 Hz; max. 2A/180 W
  • Mean consumption 80 W incl. cooler (50%): 100 W
  • Cooling capacity of compressor: 360 W (Option), max. power consumption 40W
  • Mean life cycle suction unit: > 16’000 h
  • Sensors: Ambient and flow pressure, temperature, humidity, filter storage temperature
Interfaces
  • RS232C, USB, Ethernet, RS485
  • Interface protocols:  DIGITEL, Bayern-Hessen, AK, TCP/IP, HTTP, FTP
  • Internal memory:  16MB, ring buffer, filter data of two years of daily sampling
Inlets
  • TSP, PM10, PM2.5 and PM1, customized inlets on request
Materials
  • Coated aluminium, stainless steel, POM, PTFE, NBR
  • Material of sampling line: EMATAL coated aluminium
Dimensions & Weight
  • Field housing (without inlet) 526x 235 x 1000 mm, 33 kg (36 kg with cooling), protection class IP54
  • Indoor housing (without inlet) 448 x 204 x 922 mm, 25 kg
Noise level
  • <50dB(A) at 1m, <32dB(A) at 8m
Sensor specifications
  • Flow sensor accuracy (calibrated): < 1 %
  • Ambient & internal pressure sensor range/ accuracy (calibrated): 300 – 1100 hPa/ ± 0.12 hPa
  • Ambient & internal temperature sensor range/ accuracy (calibrated): -40 – +65°C/ ± 0.5°C
  • Ambient & internal humidity sensor range/ accuracy: 0 – 60°C; 0 – 100 %/ ± 3 % rH
  • Filter storage temperature sensor range/ accuracy: -20 – +60°C/ ± 0.1°C

Options

  • Cooled filter storage with defrost/drainage system
  • Cartridges for pesticide sampling
  • LTE router for direct remote access
  • Text message (SMS) module for status and messages
  • Sampling controlled by external sensors (e.g.: wind sensor or particle counter)
  • Rotameter for auto calibration
  • Filter identification via barcode reader or RFID
  • Printer
  • High pressure drop option for membrane filter/long term sampling (800 mbar)
  • Heating for inlet or filter changer
  • Customer-specific interface protocols
  • External meteorological data collection (e.g.: wind direction and wind controlled measurement)
  • Customer- specific functions

Documents